The Importance of "Three Extractions and Three Releases" for Glove Box Entry and Exit

Glove Box For Sale

In laboratory settings and specialized work environments, glove boxes are indispensable tools. They provide a controlled, low-oxygen, or oxygen-free atmosphere essential for air-sensitive experiments and operations. When using a glove box, the process of entering and exiting items must adhere to a critical procedure known as "three extractions and three releases." This method ensures the integrity of the glove box environment, operator safety, and equipment longevity. Below, we explore the reasons behind this procedure and provide a step-by-step guide to performing it correctly.


Why Perform Three Extractions and Three Releases?

1. Maintaining a Stable Glove Box Environment

Glove boxes are designed to maintain specific gas environments, such as low-oxygen, oxygen-free, or inert atmospheres. When items are introduced or removed, outside air can enter, disrupting the controlled environment. The three extractions and three releases process effectively removes incoming air and restores the glove box to its original state. This is especially critical for experiments sensitive to oxygen and moisture, where even trace amounts can compromise results or cause adverse reactions.

2. Ensuring Operator Safety

Glove boxes often contain hazardous chemicals or gases. If the extraction and venting process is incomplete, these substances may leak when the glove box is opened, posing risks to operators. The three extractions and three releases procedure minimizes the concentration of dangerous substances, ensuring a safer working environment. It also prevents external air from reacting with hazardous materials, which could lead to dangerous situations.

3. Protecting Glove Box Equipment

Frequent opening and closing of the glove box can cause wear and tear on the equipment. Without proper extraction and venting, external air may introduce dust, moisture, or other impurities, damaging sensitive instruments. The three extractions and three releases process filters and purifies the air entering the glove box, reducing contamination and extending the equipment's lifespan.


Step-by-Step Guide to Three Extractions and Three Releases

1. First Extraction

  • Close all valves: Ensure the glove box is sealed by closing all inlet and outlet valves.
  • Start the vacuum pump: Open the exhaust valve and activate the vacuum pump to begin extracting air from the glove box.
  • Monitor pressure: Observe the pressure gauge or indicator. The extraction time depends on the glove box volume and pump power, typically ranging from a few minutes to over ten minutes.
  • Completion: When the pressure drops to the desired level, the first extraction is complete.

2. First Venting

  • Stop the vacuum pump: Close the exhaust valve and turn off the pump.
  • Introduce inert gas: Open the venting valve and slowly introduce inert gas (e.g., nitrogen) or purified air into the glove box.
  • Control venting speed: Ensure the venting is gradual to avoid sudden pressure changes.
  • Completion: When the pressure returns to near atmospheric levels, close the venting valve.

3. Second Extraction

  • Repeat extraction: Open the exhaust valve and start the vacuum pump again. This extraction can be slightly shorter than the first, as much of the air has already been removed.
  • Monitor pressure: Observe the pressure gauge until it reaches a lower level than the first extraction.
  • Completion: Close the exhaust valve and stop the pump.

4. Second Venting

  • Repeat venting: Introduce inert gas or purified air again, following the same procedure as the first venting.
  • Completion: Once the pressure stabilizes near atmospheric levels, close the venting valve.

5. Third Extraction

  • Final extraction: Perform a third extraction, which can be shorter than the previous ones. Ensure the pressure reaches the required level for your experiment or operation.
  • Completion: Close the exhaust valve and stop the vacuum pump.

6. Third Venting

  • Final venting: Introduce inert gas or purified air one last time to stabilize the pressure within the desired range.
  • Completion: The glove box environment is now fully purified and ready for safe item entry or exit.

Key Precautions During the Process

  1. Follow the steps strictly: Do not skip or alter any steps, as this could compromise the glove box environment.
  2. Check equipment functionality: Ensure the vacuum pump and valves are working properly. Repair or replace any faulty components immediately.
  3. Use high-quality inert gas: Ensure the gas or purified air meets the required standards for your experiment or operation.
  4. Monitor pressure closely: Keep an eye on the pressure gauge throughout the process. If any abnormalities occur, stop the operation and address the issue.

Jinan Kubotech co.,Ltd.

No. 887, Chunshan Road, Juyehe Street, High-tech Zone, Jinan City, Shandong Province, China.  
Email: 
vacuumdegassing@gmail.com

Phone/whatsapp/wechat/telegram: 
0086 171-8167-8326
Copyright © 2025 Vacuum Degassing Equipment Specialist. All Rights Reserved
linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram